4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
70
Order blocks
WARNING
!
Incorrect result assignment due to use of same
patient ID (if in patient mode)
The patient ID has to be unique.
r
Do not use the same patient ID for more than one
patient, otherwise results could be assigned to the
wrong patient.
Calibration and control order
q
Calibration and control orders cannot be ordered
again until the current one has completed.
Block 11: Order Deletion
Delete sample order
Order deletion is possible:
• If the order has not yet been started
• Or if the order is finished
and
all results are accepted
Delete test order
If several identical test orders exist, the first one that is
found will be deleted.
q
Ratio and profile test orders cannot be deleted via
the host interface.
Line
code
Description
Iteration
52
Special Order Selection
55
Test ID
y
Block 10: Calibration and control order
Line
code
Description
Iteration
53
Order ID
y
Block 11: Delete sample order
Line
code
Description
Iteration
(1)
(1) x = 1 to 200
53
Order ID
55
Test ID
1 to x
y
Block 11: Delete test order