4
• Glucose Control Testing
62
Starting a glucose control test
After preparing the meter as described, you can proceed
to the steps directly related to control testing:
1
From the
Main Menu
screen touch
Control Test
.
In the
Control Test
screen, the levels available for the
control solution are displayed. To the right of the buttons,
the word
Required
identifies the level at which a glucose
control test must be run to remove QC Lockout.
2
Touch
Level 1 (Lo)
or
Level 2 (Hi)
to select the level
for the following test. In the example above,
Level 2
(Hi)
is marked.
Control Test
Level 2 (Hi)
Level 1 (Lo)
Required
12:48 pm
09/18/13
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
09/18/13
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...