Initial Startup •
9
121
9.9
Options for glucose control tests
This menu contains options that allow you to specify
whether and at what intervals glucose control tests are
performed. If you require control tests, check your set-
tings for STAT tests (see page 118).
1
Touch the desired option to enable it:
■
Control Frequency
–
Always OK
: QC Lockout is disabled, patient tests
can be performed at any time. This applies regard-
less of the result of running glucose control tests.
–
Last QC OK
: QC Lockout is activated only when the
last glucose control test is out of range or a control
test is not available (e.g., when changing the test
strip lot).
–
Hours
: QC Lockout is activated after the specified
time. When a glucose control test is run success-
fully, the counter is reset to “0”.
–
Strip Count
: QC Lockout is activated after reaching
the specified number of tests. When a glucose con-
trol test is run successfully, the test strip counter is
reset to “0”.
QC Frequency
Control Frequency
Last QC OK
Always OK
Strip Count
Hours
Rotating QC
Shift
Time Of Day
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
QC Hours
50
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
QC Strips
50
12:48 pm
09/18/13
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...