Maintenance sets
The SYCON P operates largely maintenance-free. A maintenance set is available for the analyser. It
is recommended to change the hose pump cassette, hoses and O-rings after 30,000 analyses or 24
months. The unit shows a maintenance note in the display after the maintenance interval has
expired. The maintenance counter must be reset or acknowledged after maintenance has been
carried out.
Furthermore, it is recommended to clean the measuring chamber regularly, at least every 6 months.
The SYCON Clean cleaning set is offered for this purpose. It contains all the tools needed for
cleaning as well as the cleaning fluid FIT 3000
.
For information on carrying out maintenance, see the chapter "Maintenance and service".
from page 65.
Article description
Article
number
Maintenance set for SYCON 2702/2802/P
Contains the following items:
•
1 x 33-090 008 Bottle connector
•
1 x 33-090 011 Suction lance
•
1 x 33-090 217 O-ring 16x2
•
4 x 33-090 210 O-ring 9x1.5
•
1 x 33-090 218 O-ring blue
•
1 x 33-030 614 Hose pump cassette
•
1 x 33-090 002 Stirrer blade
33-090 034
Cleaning set SYCON Clean
Cleaning set for measuring chambers,
contains the following items:
•
10 pairs of gloves XL
•
1 x PP funnel
•
1 x FIT 3000 cleaner (1000 ml)
•
1 x pipette brush
•
1 x test tube brush
•
1 x container with lid
30-010 900
Measuring chamber cleaner FIT 3000 (1000
ml)
32-089 100