Maintenance and service
Good and bad counters
The good and bad counters are used to count the analysis results. They are displayed in the info
screen.
Programming: Menu > Info
Exceeding the limit value increases the bad meter, and falling below the limit value increases the
good meter. This counter can be used to assess the function of a softening system. The number of
bad measurements should be significantly smaller than the number of good measurements. The
counters can be reset in the menu.
Programming: Menu > Service > Good/Bad counter
Calibrating the unit
The unit is calibrated at the factory at a room temperature of 20 °C. When operating in very warm
or cold locations, we recommend calibrating the unit during commissioning.
To calibrate the unit, proceed as follows:
•
Carry out an analysis with the device. The p-value of the sample must be within the
measuring range of the reagent used. A measured value such as < 0.2 mmol/l cannot be
used.
•
Analyse the water in the laboratory in parallel.
•
Calculate the correction factor for the analyser using the following formula:
𝐾𝑜𝑟𝑟𝑒𝑘𝑡𝑢𝑟𝑓𝑎𝑘𝑡𝑜𝑟 =
𝑀𝑒𝑠𝑠𝑤𝑒𝑟𝑡
𝐿𝑎𝑏𝑜𝑟
𝐴𝑛𝑧𝑒𝑖𝑔𝑒𝑤𝑒𝑟𝑡
∙ 100 %
•
Set the correction factor in the unit under the specified path.
(Programming: Menu > Settings > Analysis > Calibration factor)
Example calculation:
Measured value
Labor
= 2 mmol/l
Display value from the SYCON display = 1.8 mmol/l
𝐾𝑜𝑟𝑟𝑒𝑘𝑡𝑢𝑟𝑓𝑎𝑘𝑡𝑜𝑟 =
2 𝑚𝑚𝑜𝑙/𝑙
1,8 𝑚𝑚𝑜𝑙/𝑙
∙ 100 %
𝐾𝑜𝑟𝑟𝑒𝑘𝑡𝑢𝑟𝑓𝑎𝑘𝑡𝑜𝑟 = 111,1 %
111.1 % rounded =>
correction factor = 111 %
The entry is only made with whole numbers, therefore round down/up and enter the correction
factor in the input mask on the unit.