OX-04 Operator’s Manual
Chapter 5: User Mode and Calibration
•
61
Alarm Settings (ALARM--P)
The
ALARM--P
menu has 3 items:
ALARM--P
,
DEF.ALMP
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
ALARM--P
.
2. Press and release POWER MODE. The gas name displays.
Setting the Alarm Points
1. After entering the
ALARM-P
menu, press AIR to scroll to the gas name.
2. Press and release POWER MODE.
3. The Warning setpoint flashes.
ALARM-P
O2
ALARM-P
%
AIR
AIR
AIR
O2
ALARM-P
%
DEF .ALMP
ESCAPE
O2
19.5
WARNING
%
O2
18.0
ALARM
%
O2
23.5
ALARM H
%
O2
ALARM-P
%
For all settings:
1) AIR to adjust
2) POWER MODE
to save and
continue