GD-70D
9-3
Fault Alarm Test "2-9" - "SET 4"
PW
A1
A2
F
LCD
From Detection Mode, press and hold
the MODE key for three seconds.
●
○
○
○
0.0ppm
SIH4
↓
↓
1-8.M MODE
Press the
▲
key until the Maintenance
Mode screen displays. (For GD-70D-EA
units, the Maintenance Mode screen is
item 1-9.)
●
○
○
○
1- 8
M MODE
MAINTENANCE
↓
↓
User Mode
In "1-8.M MODE", press the SET key.
●
○
○
○
1- 8
M MODE
MAINTENANCE
↓
↓
Then press and hold the SET key again
for three seconds.
●
○
○
○
- - - -
M MODE
MAINTENANCE
↓
↓
2-0.GAS TEST
The first item in Maintenance Mode will
display.
●
○
○
○
2- 0
GAS TEST
MAINTENANCE
↓
↓
2-9. SETTING1
Press the
▲
key until the
SETTING1
screen displays.
Press the SET key.
●
○
○
○
2- 9
SETTING1
MAINTENANCE
↓
↓
SET 1.ALM P
Press the
▼
key until the
F TEST
screen displays.
●
○
○
○
SET 1
F TEST
MAINTENANCE
↓
↓
SET 4.F TEST
Press the SET key.
●
○
○
○
SET 4
F TEST
MAINTENANCE
↓
↓