30 • Chapter 4: Display Mode
GasWatch 3 for CO and O
2
Last Successful Calibration Date (CAL.DATA)
The
CAL.DATA
screen shows the date of the last successful calibration. This screen only
appears if
CAL.RMDR
is set to
ON
(factory setting).
1. After entering Display Mode, press and release POWER/MODE until the
CAL.DATA
screen sequence appears.
2. Press AIR to enter the
CAL.DATA
screen and switch between the channels.
3. Press and release POWER/MODE to return to the
CAL.DATA
screen in Display Mode.
Last Successful Bump Test Date (BP.DATA)
The
BP.DATA
screen shows the date of the last successful bump test. This screen only
appears if
BP.RMDR
is set to
ON
(factory setting is
OFF
).
1. After entering Display Mode, press and release POWER/MODE until the
BP.DATA
screen sequence appears.
CAL .DATA
dISP
YES / AIR
dISP
NO / MODE
dISP
CO
1 . 1
2019
O2
1 . 15
2019
AIR
BP .DATA
dISP
YES / AIR
dISP
NO / MODE
dISP