30 • Chapter 5: Setup Mode
AirLink 9850 Gas Monitor Operator’s Manual
16. Press
MENU
.
17. 20 mA Offset: Press
ADD
or
SUB
until the device that is monitoring the 4-20 mA output reads 20
mA.
NOTE:
For a proper calibration, the 4-20 mA output must be connected to the device that is
monitoring the selected port’s 4-20 mA output.
18. Press
MENU
.
19. The AirLink 9850 returns to the 4-20 mA Output Settings menu screen.
20. Press
ADD
to continue to the TCP/IP Settings screen or press
RESET/ESC
to return to Normal
Operating Mode.
TCP/IP Settings
1. If necessary, press and hold
MENU
for 6 seconds to enter Setup Mode and then press
ADD
until
the TCP/IP Settings menu screen appears.
2. Press and release
MENU
to enter the TCP/IP Settings menu.
3. IP Address Type: Press
ADD
or
SUB
to toggle between “DHCP” and “Manual”. Contact your
network administrator to determine which option is appropriate.
4. Press
MENU
.