![RKI Instruments AirLink 7010 Operator'S Manual Download Page 25](http://html.mh-extra.com/html/rki-instruments/airlink-7010/airlink-7010_operators-manual_1466868025.webp)
AirLink 7010 Gas Monitor Operator’s Manual
Chapter 5: Setup Mode • 25
Chapter 5: Setup Mode
Overview
This mode is used for: Channel Settings (On/Off, Wired/WireFree, gas, range, decimals, radio
address), Relay Settings (On/Off, Decreasing/Increasing, Value, Latching/Auto Resetting), and System
Information.
NOTE:
Each channel must be set up individually.
NOTE:
To save any changes and exit Setup Mode at any time, press ESC. The AirLink 7010
automatically saves any changes and returns to Normal Operating Mode 15 minutes after the
last button press.
Entering Setup Mode
1. Open the enclosure box.
2. From Normal Operating Mode, press and hold
MENU,
ADD
, and
SUB
for 5 seconds to enter Setup
Mode.
Channel Settings
1. Channel Selection: Once in Setup Mode, press
ADD
(increase) or
SUB
(decrease) to select the
channel you want to set up (1-32).
NOTE:
To view system information, scroll up from channel 32 or down from channel 1.
2. Press
MENU
(Next).