35-3001A-10 Oxygen/Carbon Monoxide Sample-Draw Detector • 17
9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the
controller. RKI controllers typically have a ground stud that can be used to ground the cable’s
drain wire.
Start Up
This section describes procedures to start up the sample-draw detector and place the sample-draw
detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for
connections at the controller.
3. Turn on or plug in the power to the controller, then turn on the controller.
4. Verify that the sample-draw detector’s Pilot LED is on.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
6. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary, use the
sensor flow control valve to adjust the flow rate. Turn the valve’s knob clockwise to increase the
flow and counterclockwise to decrease the flow.
NOTE:
The following step tests for leaks in the sample line. This test will cause a low flow condition
at the sample-draw detector and a fail condition at the controller. Be sure to put the controller
into its calibration program or disable external alarms before performing this test.
7. Verify that the incoming sample line is not leaking. To test the sample line, press and hold the reset
switch and plug the open end of the sample line with your thumb. If the flowmeter ball drops to the
bottom of the flowmeter, the incoming sample line is not leaking.
8. Remove your thumb from the sample line, release the reset switch, and verify the flowmeter
returns to a normal flow rate.
9. Enable alarms or place the controller in normal operation.
CAUTION:
Allow the sample-draw detector to warm up for 5 minutes before you continue with the
next section, “Setting the Zero Reading.”
Setting the Zero Reading
CAUTION:
If you suspect the presence of carbon monoxide or an abnormal oxygen concentration in
the monitoring environment, use the calibration kit and the zero air calibration cylinder
to introduce “fresh air” to the sensor and verify an accurate zero setting. See page 25
and page 27 for instructions on using a zero air calibration cylinder for setting the zero
reading.
1. Verify that the sample-draw detector is sampling a fresh air environment (environment known to
be free of toxic gas and of normal oxygen concentration, 20.9%).
2. Open the housing door.