Setting the Confirmation Beep and
Non-Compliance Indicator (BEEP)
The
BEEP
menu has 3 items:
BEEP.SEL
,
BEEP.INT
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
BEEP
.
2. Press and release POWER MODE. The
BEEP.SEL
item appears.
BEEP.SEL
BEEP.SEL
defines what kind of confirmation or non-compliance indication you want to
occur in Measuring Mode. The available choices are:
OFF
(factory setting): The SC-04 does not provide a confirmation alert or non-compliance
indicator.
LED
: The SC-04’s LED double flashes as often as defined by the
BEEP.INT
parameter to
verify that the instrument is operating.
BUZZER
: The SC-04’s buzzer double beeps as often as defined by the
BEEP.INT
parameter
to verify that the instrument is operating.
LED+BUZ
: The SC-04’s LED double flashes and the buzzer double beeps as often as defined
by the
BEEP.INT
parameter to verify that the instrument is operating.
BMP/CAL
: If a bump test or a calibration is due and if
BP.EXPD
or
CAL.EXPD
is set to
CONFIRM
(factory setting) or
NONE
, the SC-04’s LED double flashes as often as defined
by the
BEEP.INT
parameter to indicate a non-compliance. Once a bump test or calibration
(depending on which is due) is done, the LED stop flashing.
1. While in the
BEEP
menu, press AIR to scroll to
BEEP.SEL
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BEEP.SEL
item in
the
BEEP
menu.
5. See “ESCAPE” on page 58 to return to User Mode.
BEEP
BEEP .SEL