24.3
Calibration Errors
These messages warn of incorrect calibration technique, or of attempts to calibrate the indicator beyond its
specification.
Error message
Description
ZERO BAND
The loadcell output is beyond allowable zero calibration range. Check for incorrect scale
connection. Change the dead load, or shunt the loadcells.
SPAN BAND
The loadcell signal range (span) is too small or too large for these settings. Incorrect span
weight entered. Scale wiring incorrect. Incorrect loadcell capacity. Wrong or no calibration
weight added to scale.
This error is generated if:
There are less than 100 graduations
More than 5mV/V applied
Less than -5mV/V applied
Graduation of less than 0.005uV/V per div
24.4
System Errors
The condition of the internal circuits are continuously monitored. Faults or out-of-tolerance conditions are
shown on the display as an E type error message.
Errors are added together in hexadecimal:
Example 1: If the power supply voltage is low which causes a low excitation voltage, the error message
will be
E00005
(00001
H
+ 00004
H
).
Example 2: The loadcell is not connected (i.e. both sense lines are disconnected). This will cause error
E000CO
(00080
H
+ 00040
H
).
The numbers add in hexadecimal as follows:
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F
For example:
2
H
+ 4
H
= 6
H
4
H
+ 8
H
=
C
H
A
H
+ 2
H
=
C
H
Error
Description
Action
E00001
The power supply voltage is too low
Check power supply and cables
E00002
The power supply voltage is too high
Check power supply and cables
E00004
The positive sense line is not connected
Check the loadcell connection.
This error
can be deactivated for EX-I applications with
OPTION:SENS.CH
.
E00008
The negative sense line is not connected
Check the loadcell connection.
This error
can be deactivated for EX-I applications with
OPTION:SENS.CH
.
E00010
The temperature is too hot or cold
Check the location
E00020
Accessory module failure
Restart. Replace faulty module
E00080
Resolution error
Recalibrate indicator with valid resolution
E00100
The real time clock has failed
Return for service
E00200
The setup and calibration information has been
lost
Recalibrate and reconfigure the unit
E00400
The factory information has been lost
Return for service
166
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...