6. How to Set User Mode
6-4. Alarm setpoint setting
- 46 -
6-4. Alarm setpoint setting
Set the alarm function to ON/OFF and specify an alarm setpoint.
LCD display
1
Press the ENTER
button.
The gas detector enters the
alarm setpoint setting mode.
2
Enable <on> or
disable <oFF> the
alarm function with
the
▼/▲
button.
* This is disabled <oFF> by
default.
▼
▲
3
Press the ENTER
button to confirm the
selection.
The first alarm setting screen
is displayed.
4
Set the first alarm
setpoint with the
▼/▲
button and then press
the ENTER button.
The second alarm setting
screen is displayed.
* Set the first alarm setpoint
within the following range.
Lo: 0.0 - 10.0
Hi: 11 up to the second alarm
setpoint