RM-DSA800 Installation Guide
RIGOL
1
RM-DSA800 Rackmount Kit
Product Overview
RM-DSA800, the rackmount kit of DSA800 series spectrum analyzer, as shown below, is used
to mount your analyzer into a standard 19-inch machine cabinet.
Kit Parts List
No. Name
Qty. Description
1 Front
Panel
1
2 Left
Plate
1
3
M4 Screw
14
M4 x 8 Cross Recessed Pan Head Screw
4 Spring
Washer
4
5 Washer
4
6 Support
Board
1
7 Retaining
Bracket 2
8 Right
Plate
1
9
M6 Screw
4
M6 x 16 Phil-Slot Pan Head Screw
10
M6 Nut
4
M6 x 5 Lock Blade Square Nut
3 9 10
1
2
3
4
5
6
7
8
RM-DSA800 Installation Guide
RIGOL
2
Space Requirements for Installation
The follow requirements must be fulfilled by the machine cabinet in which the instrument is
mounted.
Dimension of the machine cabinet must be standard 19-inch.
At least 5U (222.2 mm) space should be provided by the machine cabinet.
The depth inside the machine cabinet should not be less than 200 mm.
The dimension of the instrument after being installed is shown below (the unit is mm).
Installation Tool
PH2 Phillips Screwdriver (recommended)
Installation Procedures
CAUTION
Only authorized operators can execute the installation operation because
that the instrument will be damaged or installed in machine cabinet
incorrectly if the installation is not proper.
1.
Install the right and left plates: aim the detents of the right and left plates at the
openings on the support board and insert respectively, then fix them using four M4
screws.
207.5
465
220.6
133.35
403
172.2
482.6