3.1.9.8
[:SOURce][:RF]:PULM:TRAin:LIST:INSErt
Syntax
[:SOURce][:RF
<
channel>
]:PULM:TRAin:LIST:INSErt
<
OnTime>,<OffTime>,<Repeat>
Description
Inserts one line of pulse values into the pulse list for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<OnTime>
Real
20 ns to (170 s - 20 ns)
500 us
<OffTime>
Real
20 ns to (170 s -20 ns)
500 us
<Repeat>
Integer
1 to 256
1
Remarks
•
<OnTime>: sets the duration of the positive pulse.
•
<OffTime>: sets the duration of the negative pulse.
•
<Repeat>: sets the repetition times of this pulse.
Return Format
N/A
Example
:PULM:TRA:LIST:INSE 5.5s,2.5s,2 /*Inserts one line of pulse values
before the currently selected line: OnTime 5.5s, OffTime 2.5s,
Repeat 2 (by default, CH1 is selected).*/
3.1.9.9
[:SOURce][:RF]:PULM:TRAin:LIST:DELete
Syntax
[:SOURce][:RF
<
channel>
]:PULM:TRAin:LIST:DELete
<
Index>
Description
Deletes the pulse value of the specified line from the current pulse list for the
specified channel.