User’s Guide for DSA1000 Series
3.15
Amplitude Correction
Edit an amplitude-frequency curve to correct the current measured curve.
Procedures:
1.
Enter amplitude correction
-
Press AMPT
-
Page down
-
Press Corrections and enter the lower menu.
2.
Select a curve
-
Press Select
-
Press Antenna
3.
Open amplitude correction
-
Press Correction to enable or disable the amplitude correction.
-
Press Corr Table to turn on or off the correction table.
4.
Edit the correction points
-
Press Edit and enter the lower menu.
-
Press Frequency and enter 100 MHz.
-
Press Amplitude and enter 10 dB.
A correction point with 100 MHz and 10 dB is added.
-
Press Point and enter 2.
-
Press Frequency and enter 200 MHz.
-
Press Amplitude and enter 20 dB.
Another correction point with 200 MHz and 20 dB is added.