User’s Guide for DSA1000 Series
3.14
Pass/Fail Measurement
DSA1030/DSA1030-TG that has installed the Advanced Measurement Kit option
supports the pass/fail test function. In the example below, we use a generator (such
as
RIGOL
DG3101A) to output a sine signal with 100 MHz and -10 dBm as the signal
under measure.
Procedures:
1.
Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2.
Measure the signal over the Pass/Fail function
(1)
Reset the analyzer
-
Press Preset
(2)
Set the parameters
-
Press FREQ
-
Press Center Freq and enter 100 MHz.
-
Press SPAN
-
Press Span and enter 10 MHz.
-
Press Meas
-
Press Meas Fctn
—
Press the page down key
-
Press Pass/Fail
(3)
Edit the limit line
-
Press Meas Setup
-
Press Edit
-
Press Point and enter 1.
-
Press X Axis and enter 95 MHz.
-
Press Amplitude and enter -55 dBm.
-
Press Connected and select NO.
-
Add the points listed in the table below.