Table 2-33 Average number for ACP measurement
Parameter
Explanation
Default
10
Range
1 to 1000
Unit
N/A
Knob Step
1
Direction Key Step 1
2.
Avg Mode
Sets the Average operation to Exp or Repeat, the default is Exp.
In Exp mode, the result is from the exponent average calculation of the
current N (specified by Avg Num) measurements.
In Repeat mode, the result is the arithmetic average of the current N
(specified by Avg Num) measurements.
3.
Main CH BW
Sets the bandwidth of the main channel. The power of the main channel is the
power integral within this bandwidth.
You can use the numeric keys, knob or direction keys to modify this parameter.
For more details, please refer to “Parameter Input”.
Table 2-34 The main channel bandwidth of ACP measurement
Parameter
Explanation
Default
2 MHz
Range
33 Hz to 1 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Main CH BW/100, Min = 1Hz
Direction Key Step in 1, 1.5, 2, 3, 5, 7.5 sequence
4.
Adj CH BW
Sets the frequency bandwidth of the adjacent channels.
The Adj CH BW is related to the Main CH BW, which can be set within Main
CH BW/20 and Main CH BW
×
20.
You can use the numeric keys, knob or direction keys to modify this
parameter. For more details, please refer to “Parameter Input”.