Chapter 5 To Trigger the Oscilloscope
5-12
DS2000E User’s Guide
Runt Trigger
Runt trigger is used to trigger pulses that pass one trigger level but fail to pass the
other one.
For this oscilloscope, positive runt pulses are defined as pulses that pass the lower
limit of the trigger level but fail to pass the upper limit of the trigger level; negative
runt pulses are defined as pulses that pass the upper limit of the trigger level but fail
to pass the lower limit of the trigger level, as shown in the figure below.
Figure 5-4 Runt Trigger
Trigger Type:
Press Type, rotate
to select "Runt" and press down
. At this point, the trigger
setting information is displayed at the upper right corner of the screen. For example,
. The trigger type is runt trigger; the trigger source is CH1; the
difference between the upper limit of trigger level and the lower limit of trigger level
is 3.06 V.
Source Selection:
Press Source to switch the signal source list between CH1 and CH2. For the details,
please refer to the introduction in "Trigger Source". The current trigger source is
displayed at the upper right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Pulse Polarity:
Press Polarity to select the pulse polarity of runt trigger.
: positive polarity. The instrument triggers on the positive runt pulse.
: negative polarity. The instrument triggers on the negative runt pulse.
Upper Limit of the
Trigger Level
Lower Limit of the
Trigger Level
Positive Runt Pulse
Negative Runt Pulse
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...