DS2000E User’s Guide
TimeOut Trigger (Option)
In timeout trigger, the instrument triggers when the time interval (
△
T) from when
the rising edge (or falling edge) of the input signal passes through the trigger level to
when the neighbouring falling edge (or rising edge) passes through the trigger level
is greater than the timeout time set, as shown in the figure below.
Time Out
T
Time Out< T
Figure 5-10 TimeOut Trigger
Trigger Type:
Press Type to open the trigger type list. Rotate
and press down
to select
"TimeOut". At this point, the trigger setting information is displayed at the upper
right corner of the screen. For example,
. The trigger type is
timeout trigger; the trigger source is CH1; the trigger level is -1.80 V.
Source Selection:
Press Source to open the signal source list and select CH1 or CH2. For the details,
please refer to the introduction in "Trigger Source". The current trigger source is
displayed at the upper right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Edge Type:
Press Slope to select the type of the first edge of the input signal that passes
through the trigger level.
: start timing when the rising edge of the input signal passes through the
trigger level.
: start timing when the falling edge of the input signal passes through the
trigger level.
: start timing when any edge of the input signal passes through the trigger
level.
Timeout Time:
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...