Chapter 3 Constant Current Test
RIGOL
DP1116A
Performance Verification Manual
3-11
Test Record Form:
Table 3-4 CC Programming and Readback Accuracy Test Record Form
Scale
+16V/10A
+32V/5A
Specification
Programming:
0.2%* 10 mA=30
mA
Readback:
0.15%* 5 mA=20
mA
Programming:
0.2%* 10 mA=20
mA
Readback:
0.15%* 5
mA=12.5 mA
Output current is 0
DP1116A
Setting
Volt (V)
16.8
33.6
Current (A)
0
0
Measurement
Value
U
1
I
1
=U
1
/R
M
I
2
Programming
Accuracy:
|I
1
-0|
Readback
Accuracy:
|I
2
-0|
Output current is the rated output value
DP1116A
Setting
Volt (V)
16.8
33.6
Current (A)
10
5
Measurement
Value
U
3
I
3
=U
3
/R
M
I
4
Programming
Accuracy:
|I
3
-rated
output current
of the scale
under test|
Readback
Accuracy:
|I
4
-rated
output current
of the scale
under test|