5010e_513.doc / 22.01.10
7.2.13 Error
list
7.2.13 Error
list
The last 10 serious errors are shown or
printed.
The oldest error is shown first. The last error is
always marked with no 1.
By [NEXT] earlier error messages are shown.
By [PRINT] the complete error list is printed or
output to the serial interface.
For troubleshooting the coded error list can be
consulted (chapter 9.4 - CODED ERROR
MESSAGES).
E R R O R L I S T
N O . 1
D A T E 0 9 / 0 8 / 0 9
E R R O R C O D E 4 6
E X I T P R I N T N E X T
7.2.14 Key signal ON / OFF
The current status of the key signal is indi-
cated in the first line by OFF or ON.
Change setting by [SELECT].
Save setting permanently by [OK].
Save setting temporarily up to next switch-off
of the device by [EXIT].
The deeper signal tone for error messages
remains active in any case.
K E Y S I G N A L O N
E X I T O K S E L E C T
7.2.15 Touchscreen
adjustment
By this function the panel can be adjusted.
After call of the function a white cross is
shown in the left lower corner of the screen.
Touch the intersection point in the cross with a
non-scratching plastic tip (touchscreen pen,
pipette tip) as exactly as possible. In the first
line the coordinates are shown as X- and Y-
value. The input will be accepted and the
coordinate display will be reset after a time out
of 10s. Then the cross is shown in the right
upper corner. Touch the intersection point.
After a timeout of 10s the memory inquiry
follows.
Save the adjustment by [OK].
Reject the adjustment by [EXIT].
X 0 Y 0
T O U C H P A N E L A D J U S T M E N T
P R E S S E X A C T L Y T H E C R O S S P O I N T
T I M E 1 0
S A V E T P A D J U S T M E N T
E X I T O K
ROBERT RIELE GmbH & Co KG
Page 59
7- UTILITY PROGRAMS
Y PROGRAMS