30
OCP
(Over Current Protection)
Can only be enabled by Panel buttons
(
“
SHIFT
” + “
I-SET
”)
IF
the Output Current exceeds, the set value:
● Unit will enter
● The output will immediately turn OFF,
upon entering Protection Mode
.
(See Notes: **)
○ (versus the RD60xx unit adjusting output to maintain the ‘I-Set’ value.)
● The
“
ON/OFF
”
Button will turn off its illumination,
○ or react per the setting of ‘
’.
● The Current measurement value will be changed to the ‘OCP’ display colour.
○ See
‘Colours Group’ Detailed Explanation
● The
indicator is replaced by a red background white
‘OCP’ Icon.
Notes:
-
Disabled by setting to
0
.
-
When Setting,
“
SHIFT
”
button will be illuminated to visually distinguish an ‘OCP’ input
from an ‘I-SET’ input.
-
** OCP is controlled via firmware, not hardware **
.
-
As a result, there could be a delay of up to 600ms between the Current reaching
the set value and the RD60xx going into
-
Should be set to a higher value than ‘I-Set’ for proper operation.
-
See ‘
’ for an Advanced Option that allows this situation.
-
Setting OCP lower than
is typically not a good idea.
-
It was done for the example image, below, to deliberately trigger OCP.
Example Screen Image of OCP Protection activated