Page 54
Version 1.14 (2019-05)
ID-No. 9230-070-en
6.4
Switch on and off
The switching on and off of the converter
SAFERUN
® CMS-T is
done via the voltage supply. At the converter
SAFERUN
® CMS-T
there is no switch.
6.4.1
Self-test
After switching on or after a warm restart (see Section 6.4.2) or a
parameterization of the converter
SAFERUN
® CMS-T performs a
self-test. In this test, the connected sensors as well as the internal
memory as well as the parameters to be checked.
During the self-test all LEDs on the converter
SAFERUN
® CMS-T
are turned on for 5 seconds (warm start or after parameterisation
= 3 seconds).
If errors are detected during self test , this is indicated by flashing
of the corresponding LEDs.
NOTE
The chapter 8.4 Error displays contains a detailed de-
scription of the possible error codes.
i