JWI-586 DIGITAL WEIGHING INDICATOR   

9

 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

 
 
 

 

View

 

the

 

internal

 

count

 

value 

Press " PRINT/GROSS " key twice on POS 2 display. The display will show 
the internal count value. 

 

Check

 

HI.

 

OK.

 

LO

 

and

 

battery

 

signs 

Press " SET " , " PRINT/HOLD ", " GROSS/NET " and " TARE " keys in 
sequence on internal count value display. The LO, OK, HI and battery signs 
will appear in sequence accordingly. 

Then press " ZERO " key to revert to POS 2 display and press " ZERO " key 
again to return to normal weighing mode. 
 

RATIO COUNTING 

This is a simplified way for piece counting by means of weight comparison with 
the preset reference weight of certain quantity.   

1. Press " SET " key then press "GROSS/NET "until the display shows 

  " zero 

0   

pcs

 

Summary of Contents for JWI-586

Page 1: ...use troubles on the indicator apply strong solvent such as benzene and thinner for cleaning indicator It may affect on the plastic material Always operate the indicator within the temperate range from...

Page 2: ...if the backlight is always effective after the indicator is turned on When battery voltage is low a battery sign will start blinking to remind users to recharge the battery Unit will be automatically...

Page 3: ...ntainer from the platform will cause the indicator to display the container s weight as a negative number Press this key wilt return the indicator to true zero position and the ZERO sign will appear a...

Page 4: ...and OFF means DISABLE 5 Press ZERO key at any stage to revert to normal weighing mode Setup for weighing units ratio counting pcs percentage weighing and weight checking Hi Ok Lo functions 1 Press SE...

Page 5: ...1 Press TARE key the display will show FIL X FiL 2 2 Press SET key to select the approximate level for your application and leave the digital filtering setup Each level is a trade off between reactio...

Page 6: ...ent enabled setting for this setup is lit Aut Zero band setup Input the weight range where the unit will display zero The weight range is in terms of number of display divisions set for both positive...

Page 7: ...his setup is Prt Pr Hold time setup This function enables the user to hold the displayed stable readings for a pre set time period of none 5 10 20 or 40 seconds 1 Press TARE key the display will show...

Page 8: ...key then press SET key The display will show 6000 the pre shipment setting for capacity 4 Press TARE key to revolve the allowed capacities as per the following cap div table until the desired capacit...

Page 9: ...ys in sequence on internal count value display The LO OK HI and battery signs will appear in sequence accordingly Then press ZERO key to revert to POS 2 display and press ZERO key again to return to n...

Page 10: ...your application 5 Put on the objects intended for piece counting the calculated quantity will be displayed PERCENTAGE WEIGHING This function can be used for user defined weighing unit or material mi...

Page 11: ...he high limit weight into memory and then return the unit to normal weighing mode 4 Press SET key and then press PRINT HOLD key The display will show 5 Repeat step 2 to determine the low limit of the...

Page 12: ...ed up Otherwise the indicator will lock a new greater weight value 1 Press SET key and then press ZERO key The display will show 2 The unit at this time is ready for instant max Impact lockup PRE SET...

Page 13: ...CATOR 13 OPTIONAL RS 232 SERIAL PORT 25 PIN JUNCTION BOX 1 4 OP 01 RS 232C 1 Output data Baud rate 2400 4800 or 9600 bps Data bit 7 Parity bit E even Stop bit 1 Code ASCIl Bit format LSB MSB DATA BIT...

Page 14: ...0 GOTO 20 50 END 2 Input commands or preset tare high limit low limit values from computer a Input commands R READ weight data T perform TARE function N view NET weight Z perform ZERO function b Prese...

Page 15: ...t digits after decimal is 1 L004800 preset low limit value at 48 if the previously preset digits after decimal is 2 Example of programs in BASICA for executing commands 10 OPEN COM1 9600 E 7 1 RS DS L...

Page 16: ...s great as possible A Load cell rated load B Load cell rated output mV V D Minimum division 5000 is the excitation voltage in millivolts 5V 0 6 is the input sensitivity in V A D B 5000 6 0 Example To...

Reviews: