![Rice Lake TA Plus 124 Operation Manual Download Page 66](http://html1.mh-extra.com/html/rice-lake/ta-plus-124/ta-plus-124_operation-manual_1454254066.webp)
60
Visit our website
TA Plus Balance
7.5 Command Formats
Do not confuse the alphabetical O for Arabic number 0 (Zero).
C1
C2
Code
(
C1)
Code
(
C2)
Description
Response
A00/Exx
format
ACK/NAK
format
T
(SP)
0x54
0x20
Tare
A00:
Normal
response
E01:
Abnormal
response
ACK:
Normal
response
NAK:
Abnormal
response
Z
(SP)
0x5a
0x20
Zero-point adjustment
O
0
0x4f
0x30
Stop output
O
1
0x4f
0x31
Continuous output
O
2
0x4f
0x32
Continuous output (no output when unstable)
O
3
0x4f
0x33
Press output key for one-time output
O
4
0x4f
0x34
Auto output
O
5
0x4f
0x35
One-time output when stable
O
6
0x4f
0x36
One-time output when stable
O
7
0x4f
0x37
Press output key for one-time output when stable
Commands O0 to O7:
• Have the same functions as the output control set by the setting menu
• Once commands are executed, that state is maintained; the status is reset to the setting menu when the balance is
turned on again
O
8
0x4f
0x38
One-time output
O
9
0x4f
0x39
One-time output after stability is obtained
Command O8 to O9:
• Are used to request data from the balance
• After the command is executed, it returns to O0
O
A
0x4f
0x41
Interval function (Output each time the output time has elapsed)
O
B
0x4f
0x42
Interval function (Output when stable each time the output time
has elapsed)
When OA or OB command is sent, the interval function starts and must be sent again to end the function
Table 7-6. Zero Point Adjustment/Tare/Output Control Setting Command
C1
C2
Code (C1)
Code (C2)
Description
Response
D
D
0x44
0x44
Date output request
Date data
D
T
0x44
0x54
Time output request
Time data
Table 7-7. Data Output Request and Time Output Request