Rice Lake Weighing Systems
●
All Rights Reserved
27
Span Test and Calibration
4. Press
to select the
NTF
setting. Options available are:
4.2.6 Weight Error Input
The error of the external reference weight used for span adjustment/span test can be inputted.
1. Press
to change the mode.
2. Press
to scroll to
Maintenance
.
3. Press
to scroll to
639 ERROR INPUT
.
4. Input the unit weight by pressing Press
through
.
5. Press
to save the selection.
6. Press
and
63A ERROR ON
displays.
7. Press
then
to select
ON
.
8. Press
to save the selection.
9. Press
to exit the setting menu.
10. Perform and span adjustment/test weigh external weights (
).
4.2.7 Weight Error Reflection
The error of the external reference weight used for span adjustment/span test can be inputted.
1. Press
to change the mode.
2. Press
to scroll to Maintenance.
3. Press
to scroll to
63A ERROR INPUT
.
4. Input the unit weight by pressing Press
through
.
5. Press
to save the selection.
OFF
Disable
NTF
Advice CAL
AUTO Auto CAL
Table 4-2. Alarm Calibration Settings
5. Press
to change the mode.
6. Press
to exit the setting menu.
7. A message appears (
PLEASE EXEC CAL RET
) when time passed since a power up, a last span adjustment or
temperature pressure changed.
8. Press
to shift to the measuring mode.
9. Start a span adjustment with external weights (
) or a semi-automatic span adjustment with
internal weights (
).