Introduction
5
1.3
Quick Setup Menu
Press
to turn the instrument on. Enter the quick setup menu by pressing
as the firmware version displays.
Figure 1-3. Quick Setup Menu
1.3.1
Default Factory Calibration
The instrument is shipped with the following default calibration settings:
• Capacity – 10,000 kg
• Load cell sensitivity – 2.000 mV/V
• Division – 1
To default the instrument to factory settings use the following steps.
1. Turn on the instrument and press
while the firmware version displays.
displays.
2. Press
or
until
displays.
3. Press .
displays.
C
Press and to set decimals and minimum division.
Press , and to set the total capacity of the load cells.
Press , and to set the mV/V sensitivity of the load cell.
Press , and to set the weight value present on the load cell.
Press , and to set the scale capacity.
Quick calibration of zero.
Press , and to set the weight value used for the calibration.
Check the load cell signal in millivolts.
Set the analog output on net or gross weight.
Set the minimum value of the analog output.
Set the analog output when the scale shows zero weight.
Set the maximum value of the analog output.
Set the key to be emulated by the input.
Set the output normal state.
Set the output switching condition.
Enable time to disable the output.
Enable time to enable the output.
Enable or disable the hysteresis.
= Optional Analog Model
When settings are complete press until the indicator displays
SAVE?
. Press to save set up and return to the weigh mode.
Pressing any other key exits the setup and discards changes.
INPUTS
CLOCK
0 NONE
1 GROS
2 NET
3 PCS
4 GRO.0
5 NET.0
6 MOTI
23 K.PR
25.K.MOD
26 K.C
27.K.ZER
28.K.TAR
29 ERR
DIV.DEC
CEL.CAP
CEL.SEN
DEAD.LD
RANGE
ZERO
WEIGHT
ADC.MVV
AN.OUT
AO ZER
MODE
AO MIN
INP.02
AO MAX
INP.01
INP.02
OUTPUT
Disabled.
Set point on the gross weight.
Set point on the net weight.
Set point on the pieces.
Gross weight at zero.
Net weight at zero.
Active on instability.
Active when the
C
key is pressed.
Error indication.
Active when is pressed.
Active when is pressed.
Active when is pressed.
Active when is pressed.
NO/NC
ONSTAT
RL.IST
ENAB.TM
DELAY
INP.01
FUNC
1
2
3
4
5
6
7
8
9
10
11
12