30
SCT-1100 Weight Transmitter
Manual Entry of APW
A known APW can be entered with the keyboard.
To display the stored APW:
• Press and hold
.
and the stored nine digit value with five decimal places displays in the
programmed unit of measure.
To enter a known APW:
1. Press to clear the present value.
2. Press
to select the digit to be modified and
or
to increase or decrease the digit.
3. Press
to
confirm.
Example: Unit of measure of the APW in g
0000.00000 is 0.0 g, therefore by entering 0001.05000 the weight is 1.05 g.
Printing
Press
to print the data as it was programmed in the
of the setup mode
).
For example: gross weight, tare weight, net weight, quantity of pieces (PCS) currently on
the scale, calculated APW.
3.10.13Simultaneous Transmitter Mode
The independent channel mode makes it possible to have simultaneous transmission of the values on each single
channel via the serial line. In addition, through the optional alibi memory board, the transmitted weight values can
be stored in a computer for data processing and/or integration. The filed values can then be recalled from the PC
serial line or directly on the instrument’s display.
In this mode the zero tracking and the scale keys
,
and
are disabled. It is not possible to set
functioning modes and the alibi mode is set automatically.
To set the transmitter mode:
1. Turn on the instrument and press
while the firmware version displays.
displays.
2. Press
to enter the menu.
3. Press
or
until
displays. Press
to select.
4. Slowly press multiple times until
displays. Press
to confirm.
Operation
In the simultaneous transmitter mode it is possible to view the weighing information. See
Reviewing Stored Weigh Information
• Press
to switch channels, if the scale is configured as multichannel.
• Only functions which can be enabled through serial commands can be performed. See
regarding
Serial Commands in Alibi Memory Mode
. Not all serial commands can be performed.
. It is not possible to carry out other operations in the simultaneous transmitter
mode.
• The list of usable serial commands includes:
PID, ALRD, ALDL, VER, REXT, REXTA, READ, MVOL, RAZF,
CGCHN, ECHO, DISP, DINT, PCOK, STAT, KEYP, KEYR, KEYEE,KEYED
Pressing simulation of the scale keys through the KEYP and KEYR commands allows for management of the
functions linked to the
key.