RL-350 Series Wheelchair Scales
18
Visit our website
13. Press
again and the display indicates
Done
.
14. Press
three times to exit back out to the top level
Done
parameter.
15. Press
to return to
Weigh
mode.
16. To exit calibration without changing zero or span existing calibration, press
, then press
.
4.7 Test Menu
To access the
TEST
menu, use the following steps.
1. Press
and
simultaneously to power on the scale until
ID
flashes.
2. Press
again.
3. Continue to press
to scroll through the various menu items.
4. Once complete, press
again then
Done
displays.
5. Press
to start the weighing process.
Figure 4-5. Test Menu
Parameter
Choice
Description
RID
Value
Displays internal ID number
VER
Value
Displays the current software version
BAT
Value
Displays the current battery level
VALUE
Value
Displays the actual value
A2D
Value
Displays the actual raw counts of the scale
Table 4-2. Test Menu
ENTER
BMI
ENTER
CLEAR
BMI
ENTER
ENTER
BMI
BMI
ENTER
DEF
CALIB
PROG
DONE
TEST
VER
VALUE
A2D
Value
Value
Value
Value
BAT
RID