Fieldbus Options Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
77
10.5.1 Commands
Table 10-4. Command Values
Command
Decimal
Description
No Command
0
No action
Zero Scale
1
Zero the scale
Tare Scale
2
Tare the current weight if parameter 1=0, otherwise tare
the value that is in parameter 1
Clear Tare
3
Clear a tare if one exists
Display Net Mode
4
Switch the scale from gross mode to net mode
Display Gross Mode
5
Switch the scale from net mode to gross mode
Write Setpoint
10
Parameter 1 = setpoint # parameter 2 = value
Read Setpoint
11
Parameter 1 = setpoint #
Read IO Points
12
Parameter 1 = IO slot number
Set Output On
24
Parameter 1 = slot parameter 2 = bit
Set Output Off
25
Parameter 1 = slot parameter 2 = bit
Write Setup
27
Write calibration settings and enter calibration mode
Restart Instrument
34
Restarts the indicator
Perform Zero Calibration
35
Performs a zero calibration
Perform Span Calibration
36
Performs a span calibration
Point Calibration
37
Point value found in parameter 1 (1-3)
Save Calibration
38
Save the calibration and exit calibration mode
Abort Calibration
39
Aborts a calibration and clears any errors
Keyboard Enable/Disable
40
Disable keys (parameter 1 = 0)
Enable keys (parameter 1 = 1)
Read Accumulator
41
Multivalue 1 = returned accumulator value