680 Synergy Plus – Digital Weight Indicator
40
Visit our website
5.0
Calibration
The 680 can be calibrated using the front panel and EDP commands. The following sections describe the procedures required
for these calibration methods.
The 680 requires a WZERO and WSPAN points to be calibrated. The linear calibration points are optional; they must
fall between zero and span, but must not duplicate zero or span.
Figure 5-1. Calibration Menu
5.1 Front Panel Calibration
5.1.1 Span Calibration
Use the following steps to perform a standard span calibration on a connected scale.
1. Access the setup menu by pressing the setup switch (
).
Config
displays.
2. Press
twice.
Calibr
displays.
3. Press
.
wzero
displays.
See
if the application requires a rezero, last zero or temporary zero.
4. Ensure there is no weight on the scale.
5. Press
to perform a zero calibration.
ok
displays.
6. Press
.
zerocnt
zerocnt
.
7. Press
.
wval
displays.
8. Press
. The current test weight value displays.
9. Press
and use the numeric keypad to enter a new value, if necessary.
10. Press
to accept value.
wSpan
displays.
11. Place the specified amount of test weight on the scale.
12. Press
to perform a span calibration.
ok
displays.
13. Press
.
spancnt
displays. See
zpancnt
.
Span calibration is complete. To continue with a linear calibration, see
before returning to
weigh mode.
14. Press
to return to weigh mode.
Note
Setup
Calibr
zerocnt
WZERO
WVAL
spancnt
WSPAN
Point 1-4
WLIN
REZERO
Wlin v1-4
wlin c1-4
wlin f1-4
lstzero
tmpzero
Note
Note