Calibration
29
4. Enter the
Value of Test Weight
to be used for span calibration then click
OK
.
5. The Zero Calibration dialog box prompts you to remove all weight from the scale. Clear the scale and click
OK
to begin zero calibration.
NOTE:
If your test weights require hooks or chains, place the hooks or chains on the
scale for zero calibration.
6. When zero calibration is complete, the Span Calibration dialog box prompts you to place test weights on
the scale for span calibration. Place tests weights on the scale then click
OK
.
7. The Rezero dialog box prompts you to remove all weight, from the scale (including hooks or chains).
Remove the weights then click
OK
.
8. When calibration is complete, the
New Settings
fields of the Weight Calibration display are filled in. Click
Exit
to save the new values and return to the Revolution main menu; to restore the previous calibration
values, click
Restore Settings
.
Figure 4-3. Revolution Calibration Display