3.2 Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the
IQ plus 210
menu structures. In the actual menu structure, the
values under each parameter are arranged horizontally. To save page space, menu choices are shown in vertical columns. The
factory default setting appears at the top of each column. Most menu diagrams are accompanied by a table that describes all
parameters and parameter values associated with that menu. Default parameter values are shown in bold type.
3.2.1 Configuration Menu
Figure 3-6. Configuration Menu
CONFIG Menu
Parameter
Choices
Description
Level 2 submenus
GRADS
5000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in the range 1–10 000
and should be consistent with legal requirements and environmental limits on system resolution.
To calculate GRADS, use the formula, GRADS =
Capacity
/
Display Divisions
.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBN
OFF
0.5D
1D
3D
Zero track band. Automatically zeroes the scale when within the range specified, as long as the input is within
the configured zero range (ZRANGE parameter). Selections are ± display divisions. Maximum legal value var
-
ies depending on local regulations.
ZRANGE
100%
1.9%
Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is ± 1.9% around the
calibrated zero point, for a total range of 3.8%. Indicator must be at standstill to zero the scale. Use 1.9% for
legal-for-trade applications.
MOTBAN
1D
2D
3D
5D
10D
20D
OFF
Motion band. Sets the level, in display divisions, at which scale motion is detected. If motion is not detected for
1 second or more, the standstill symbol lights. Some operations, including print, tare, and zero, require the
scale to be at standstill. Maximum legal value varies depending on local regulations.
If OFF is selected, ZTRKBN should also be set to OFF.
OVRLOA
FS+2%
FS+1D
FS+9D
FS
Overload. Determines the point at which the display blanks and an out-of-range error message is displayed.
Maximum legal value varies depending on local regulations.
FILTER
4
8
16
4RT
8RT
16RT
1
2
Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical vibration from the
immediate area of the scale.
Choices indicate the number of A/D conversions that are averaged to obtain the displayed reading. A higher
number gives a more accurate display by minimizing the effect of a few noisy readings, but slows down the
settling rate of the indicator. Values with the RT suffix provide a raised threshold for filtering larger noise
spikes.
Table 3-2. Configuration Menu Parameters
XXXXXXX
XXXXXXX
GRADS
5000
OFF
1D
0.5D
3D
ZTRKBN
ZRANGE
100%
1.9%
OFF
2D
1D
3D
MOTBAN
10D
5D
20D
FS+2%
FS+9D
FS+1D
FS
OVRLOA
1
8
4
16
FILTER
8RT
4RT
16RT
number
VERS
DEFLT
DIG IN
SERIAL
CALIBR
CONFIG
FORMAT
2