![Rice Lake 380 Synergy Technical Manual Download Page 44](http://html1.mh-extra.com/html/rice-lake/380-synergy/380-synergy_technical-manual_1453824044.webp)
380 Synergy Digital Weight Indicator
38
Visit our website
7.6 EDP Setting Commands
7.7 Stream Setting Commands
7.8 Feature Commands
Command
Description
Values
EDP.TRIGGER#
p
Port serial function
CMD
(default), STRIND, STRLFT
EDP.BAUD#
p
Port baud rate
1200, 2400, 4800,
9600
(default), 19200, 38400, 57600, 115200
EDP.BITS#
p
Port data bits/parity
8NONE
(default), 7EVEN, 7ODD
EDP.TERMIN#
p
Port line termination characters
CR/LF
(default), CR
EDP.STOPBITS#
p
Port stop bits
1
(default), 2
EDP.ECHO#
p
Port echo
ON
(default), OFF
EDP.RESPONSE#
p
Port response
ON
(default), OFF
EDP.EOLDLY#
p
Port end-of-line delay
0–255 (0.1-second intervals),
0
(default)
For commands ending with
#p
, p is the port number (1); This is a 1-port device
Table 7-6. Serial Port Commands
Command
Description
Values
STRM.FORMAT#
n
Stream format
RLWS
(default), CRDNAL, WTRNIX, TOLEDO
For commands ending with
#n
, n is the stream format number (1)
Table 7-7. Stream Formatting Commands
Command
Description
Values
DATEFMT
Date format **
MMDDYY
(default), DDMMYY, YYMMDD, YYDDMM
DATESEP
Date separator **
SLASH
(default), DASH, SEMI, DOT
TIMEFMT
Time format **
12HOUR
(default), 24HOUR
TIMESEP
Time separator **
COLON
(default), COMMA, DOT
BRIGHTNESS
Brightness of the backlight
OFF,
LOW
(default), MED, HIGH
AUTOFFTIME
Auto off time
0-60 seconds,
0
=Always On (default)
BATTERYINSTALLED
Power source
NONE,
ALK
(default)
POWERSAVEMODE
Power save mode
NONE
(default), LOW, HIGH
CONSNUM
Consecutive numbering
0–999999,
0
(default)
CONSTUP
Consecutive number start-up value
0–999999,
0
(default)
UID
ID of the indicator
Up to 8 alphanumeric characters,
1
(default)
KYBDLK
Keyboard lock (disable keypad except power)
OFF
(default), ON
ZERONLY
Disable all keys except ZERO and power
OFF
(default), ON
KHOLDTIME
Key hold time (in tenths of a second); 20 equals 2 seconds
10–50,
20
(default)
KHOLDINTERVAL
Key hold time interval; the amount of time between increments
during a key hold (in twentieths of a second); 2 equals a tenth
of a second (10 increments per second during a key hold)
1–100,
2
(default)
LOCALE
Enable gravity compensation
OFF
(default), FACTOR
GRAV.LOC
Origin gravity factor (in m/s
2
) for gravity compensation
9.00000–9.99999,
9.80665
(default)
DEST.GRAV.LOC
Destination gravity factor (in m/s
2
) for gravity compensation
9.00000–9.99999,
9.80665
(default)
** Date/time functions only available with RTC option
Table 7-8. Feature Commands