EDP Commands
© Rice Lake Weighing Systems
●
All Rights Reserved
37
7.5 USB Setting Commands
SC.MAX_DATE#n
Date of maximum weighment (read only)
Up to 25 alphanumeric characters (Only available if RTC option
is installed)
SC.WMTTHRH#n
Weighment threshold
0.0–999999.0,
1000.0
(default)
SC.DIGFLTR1#
n
SC.DIGFLTR2#
n
SC.DIGFLTR3#
n
Number of A/D samples averaged for the individual
stages (1-3) of the three stage digital filter
1, 2,
4
(default), 8, 16, 32, 64, 128, 256
SC.DFSENS#
n
Digital filter cutout sensitivity
2OUT
(default), 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
SC.DFTHRH#
n
Digital filter cutout threshold
NONE
(default), 2D, 5D, 10D, 20D, 50D, 100D, 200D, 250D
SC.SMPRAT#
n
Scale A/D sample rate
5HZ, 6HZ, 10HZ, 12HZ, 20HZ,
25HZ
(default), 40HZ, 50HZ,
80HZ, 100HZ
SC.PWRUPMD#
n
Power up mode
GO
(default), DELAY
SC.TAREFN#
n
Tare function
BOTH
(default), KEYED, NOTARE, PBTARE
SC.PRI.FMT#
n
Primary units format (decimal point and
display divisions)
888100, 888200, 888500, 888810, 888820, 888850,
888881
(default), 888882, 888885, 88888.1, 88888.2, 88888.5,
8888.81, 8888.82, 8888.85, 888.881, 888.882, 888.885,
88.8881, 88.8882, 88.8885, 8.88881, 8.88882, 8.88885
SC.PRI.UNITS#
n
Primary units
LB
(default), KG, OZ, TN, T, G, NONE
SC.SEC.FMT#
n
Secondary units format (decimal point and
display divisions)
888100, 888200, 888500, 888810, 888820, 888850, 888881,
888882, 888885, 88888.1, 88888.2,
88888.5
(default), 8888.81,
8888.82, 8888.85, 888.881, 888.882, 888.885, 88.8881,
88.8882, 88.8885, 8.88881, 8.88882, 8.88885
SC.SEC.UNITS#
n
Secondary units
LB,
KG
(default), OZ, TN, T, G, NONE
SC.SEC.ENABLED#
n
Enable the Secondary Units
ON
(default), OFF
SC.FILTERCHAIN#
n
Defines what filter to use
AVGONLY
(default), RAW
SC.RTZGRAD#
n
Number of graduations from the zero base at which the
accumulator re-arms itself
0.0–100.0,
0.4
(default)
SC.ACCUM#
n
Accumulator enable
OFF
(default), ON
SC.WZERO#
n
Perform zero calibration
—
SC.TEMPZERO
#n
Perform temporary zero calibration
—
SC.LASTZERO
#n
Perform last zero calibration
—
SC.WVAL#
n
Test weight value
0.000001–999999.999999,
10000.0
(default)
SC.WSPAN#
n
Perform span calibration
—
SC.WLIN.F1#
n
–
SC.WLIN.F4#
n
Actual raw count value for linearization points 1–4
0–16777215,
0
(default)
SC.WLIN.V1#
n
–
SC.WLIN.V4#
n
Test weight value for linearization points 1–4 (A setting
of 0 indicates the linearization point is not used)
0.000001–999999.999999,
0.0
(default)
SC.WLIN.C1#
n
–
SC.WLIN.C4#
n
Perform linearization calibration on points 1–4
—
SC.LC.CD#
n
Zero (deadload) coefficient raw count value
0–16777215,
8386509
(default)
SC.LC.CW#
n
Span coefficient raw count value
0–16777215,
2186044
(default)
SC.LC.CZ#
n
Temporary zero raw count value
0–16777215,
2186044
(default)
SC.REZERO#
n
Perform the Rezero calibration function
—
Command
Description
Values
USB.TRIGGER
USB serial function
CMD
(default), STRIND, STRLFT
USB.LINETERM
USB line termination characters
CR/LF
(default), CR
USB.ECHO
USB echo
ON,
OFF
(default)
USB.RESPONSE
USB response
ON
(default), OFF
USB.EOLDLY
USB end-of-line delay
0–255 (0.1-second intervals),
0
(default)
Table 7-5. USB Port Commands
Command
Description
Values
For commands ending with
#n
, n is the scale number (1)
Table 7-4. Scales Commands (Continued)