Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
27
4.3.5 Setup – Program Menu
Figure 4-10. Setup – Program Menu
** Only visible with RTC option installed.
Parameter
Description
pwrupm
Power Up Mode – When the indicator is turned on, it performs a display test, then enters a warm up period;
Settings:
GO
(default)
– Performs display test, then enters weigh mode after brief warm up period
DELAY
– Performs display test, then enters a warm up period of 30 seconds
• If no motion is detected during warm up period, indicator enters weigh mode when warm up period ends
• If motion is detected, the 30 second timer is reset and the warm up period is repeated
regula
Regulatory Mode – Specifies the regulatory agency having jurisdiction over the scale site; the value specified for this parameter
affects the function of the front panel tare and zero keys; Settings:
NTEP
(default), OIML, CANADA, NONE
• OIML, NTEP, and CANADA modes allow a tare to be acquired at a weight greater than zero; NONE allows tares to be
acquired at any weight value
• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no load; NONE allows tares to be
cleared at any weight value
• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present; In CANADA mode, the previous
tare must be cleared before a new tare can be acquired
• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as long as the current weight is
within the specified ZRANGE; In OIML mode, the scale must be in gross mode before it can be zeroed; pressing the ZERO
key in net mode clears the tare
consnu
Consecutive Numbering – Allows sequential numbering for print operations; the value is incremented following each print operation
which includes <CN> in the ticket format;
Enter value:
0–999999,
0
(default)
constu
Consecutive Number Startup Value – Specifies the initial consecutive number (CONSNU) value used when the consecutive number
is reset by sending KCLRCN EDP serial command;
Enter value:
0–999999,
0
(default)
uid
Unit ID – Specifies the unit identification string with an alphanumeric value;
Enter characters: Up to 8 alphanumeric characters,
1
(default)
accum
Accumulator – Accumulation can be toggled ON/OFF; if ON, accumulation occurs on print operation; if OFF, an accumulation
does not occur;
Settings:
OFF
(default), ON
rtZGrd
Return to Zero Grads – When the weight (in display graduations) falls below the value set, the accumulator is rearmed;
Enter Value: 0.0-100.0,
0.4
(default)
WMTTHR
Weighment Threshold – When the weight exceeds the value set, the number of weighments counter is incremented;
Enter value: 0.0–999999.0,
1000.0
(default)
NUMWGH
Number of Weighments – Displays the total number of weighments;
Read only
MAXWGH
Maximum Weighment – Displays the maximum weighment made;
Read only
MAXDAT
Date/Time of Max Weight – Displays the date the maximum weighment occurred;
Read only (only available when RTC is installed)
date
Date – Allows setting of the date format and date separator character;
(only available when RTC is installed)
DATEFMT – Date Format; Settings:
MMDDYY
(default), DDMMYY, YYMMDD, YYDDMM
DATESEP – Date Separator; Settings:
SLASH
(default), DASH, SEMI, DOT
time
Time – Allows setting of the time format and the separator character;
(only available when RTC is installed)
TIMEFMT – Time Format; Settings:
12HOUR
(default), 24HOUR
TIMESEP – Time Separator; Settings:
COLON
(default), COMMA, DOT
Table 4-8. Setup – Program Menu Description
Setup
Progrm
Pwrupm
Regula
Consnu
Uid
constu
date
time
rtzgrd
wmtthr
maxwgh
kybdlk
zeronl
bright
numwgh
Accum
Locale
grvorg
grvdst
keyhld
autoff
Batery
pwrsav
maxdat
Displays if
ACCUM
is ON
Displays if
locale
is set to
Factor
**
**
**
Note