background image

Model number: UT1033

PLEASE READ THE USER’S 

MANUAL CAREFULLY BEFORE

USING THIS PRODUCT

USE ONLY AS DIRECTED

If symptoms persist, consult  

your healthcare professional

User’s Manual

3269_IFU02_14269060.indd   1

24/07/2017   11:59

Summary of Contents for UT1033

Page 1: ...r UT1033 PLEASE READ THE USER S MANUAL CAREFULLY BEFORE USING THIS PRODUCT USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional User s Manual 3269_IFU02_14269060 indd 1 24 07...

Page 2: ...gs and Cautions 8 Instructions for use 9 16 How to use REVITIVE Ultrasound 9 11 Therapy Application Table 12 13 Treatment Head Test 14 15 How to use the Ultrasound Gel 15 Cleaning Recommendations 16 S...

Page 3: ...opening the carton please check that the following components are provided If you think anything is missing please contact us using the helpline numbers on the back of this booklet What is inside the...

Page 4: ...ll or reproductive organs Used over open wounds or fragile or damaged skin eg eczema Used on the front of the neck over the carotid sinus Used over spinal abnormalities e g spina bifida following lami...

Page 5: ...and consult your health care professional The socket outlet should be installed near the equipment and should be easily accessible Use only the accessories supplied by or purchased from the manufactu...

Page 6: ...e Time Indicators When in use the chosen Time Indicator will be on constantly and the treatment will count down for the chosen Time duration If the Time Indicator is flashing there is not enough press...

Page 7: ...inutes per treatment head area 1 0 4 LOW 4 minutes per treatment head area 2 0 8 LOW 4 5 minutes per treatment head area 3 1 2 MEDIUM 2 minutes per treatment head area 4 1 6 MEDIUM 2 5 minutes per tre...

Page 8: ...ater will visibly vibrate demonstrating that the Treatment Head is working correctly Important The Treatment Head only vibrates if the conductive fluid water or gel covers the entire surface of the Tr...

Page 9: ...Pa 700 hPa 1060 hPa No vibration Ultrasound waves with a vibra tion of 1 million times per sec ond are too fast to see or feel Therefore you will not feel any sensation when using REVITIVE Ultrasound...

Page 10: ...rical Device Type BF IEC 60601 1 6 1l Serial number is on the underside of the device and on the packaging Directive 93 42EEC Annex 1 13 3 d EN 15223 1 2012 Lot Number including year YY and Month MM o...

Page 11: ...ICATIONS EQUIPMENT AND EQUIPMENT OR SYSTEMS THAT ARE NOT LIFE SUPPORTING Technical Specifications Technical Specifications REVITIVE Ultrasound is intended for use in the electromagnetic environment sp...

Page 12: ...l be calculated from the original purchase date Exclusions Actegy Ltd shall not be liable to replace or repair the goods under the terms of the guarantee where Damage or the fault is due to improper u...

Page 13: ...management standard for medical devices Made in PRC Shanghai International Holding Corp GmbH Europe Eiffestra e 80 20537 Hamburg Germany Shenzhen Dongdixin Technology Co Ltd No3 Building Xilibaimang...