background image

                                                                                          

 

 
 

EN   Instruction manual 
 

 

 

Baignoire à Hauteur Variable avec Lève-personne 
Variable Height Bath and Integral Transfer System 

 
Caprice 

 
7101.01 UK 
7101.02 UK 

 

 

 

Summary of Contents for Caprice 7101.01 UK

Page 1: ...EN Instruction manual Baignoire Hauteur Variable avec L ve personne Variable Height Bath and Integral Transfer System Caprice 7101 01 UK 7101 02 UK...

Page 2: ...8 SEATING SYSTEM AND HOIST INTERFACE 8 RAISING THE SEATING SYSTEM 9 LOWERING THE SEATING SYSTEM 9 AUTOMATIC SEQUENCES FUNCTIONS 9 TEMPERATURE CONTROL 10 DRAIN PLUG OPERATION 11 HYDRO MASSAGE OPERATION...

Page 3: ...DIMENSIONS 2 18...

Page 4: ...omatic transfer to trolley Sequence A 1 Tub fully comes down 2 Hoist fully rises 3 Seat fully turns to transfer Automatic transfer to bath Sequence B 1 Hoist fully rises 2 Tub fully comes down 3 Seat...

Page 5: ...Filling time 3 bars 4 min Safe working load SWL 125Kg Working capacity for Hi Lo tub unlimited 6 min hours Working capacity for hoist unlimited 6 min hours ELECTRICITY Electric supply requirements 23...

Page 6: ...reasons the seating system will only rotate when the tub is in LOW position and the hoist in HIGH position To raise the tub is only possible if the seating system is ABOVE TUB LEVEL Lowering the seat...

Page 7: ...D DO NOT immerse the hand control in water DO NOT try to lift the tub DO NOT work underneath the tub during lowering DO NOT fill over the maximum water level line DO NOT operate the air whirlpool pump...

Page 8: ...initialise in end stop inwards The initialisation procedure is as follows 1 Press on the button to endstop sound signal then 2 Press on the button to endstop sound signal then 3 Press on the button to...

Page 9: ...ould take special care not exert excess force or exceed a slow constant and controlled walking pace Care should also be taken when crossing room thresholds or very small obstacles which may require th...

Page 10: ...g system to go out the bath press the ROTATE button until endstop sound signal or to go in the bath BATH HEIGHT ADJUSTMENT Raise the bath tub to the required height then lower the seating system into...

Page 11: ...ital mixer Close the drain plug by turning the button in a clockwise direction Pass the finger in front of the pictogram of the tap PATIENT SHOWER The shower head can be operated at any time Pass the...

Page 12: ...The three button control is located on the tub System Safety Drying cycle auto switch off If the hydro massage function is used at any time 30 minutes after use the hydro massage bower automatically...

Page 13: ...sation This coupled with Reval s Hydro Aromatherapy system deliver the ultimate in bathing comfort relaxation and rehabilitation The Chromotherapy control button allows the user or care giver total co...

Page 14: ...matherapy system on via the circular control panel Once up to temperature the heated air infuses with the aroma oil and releases its fragrance into the water via the diffuser jets positioned in the bo...

Page 15: ...tion Take the MP3 player Set it and add in musics of your choice Respect the legislation Connect the MP3 player headphone plug to the DIN plug located inside the embeded black box Read your MP3 player...

Page 16: ...er litre of cold fresh water Health and Safety When mixing or applying REVAL disinfectant always ware protective rubber gloves safety clothing and appropriate eye protection Method For 2 4 mltrs of ne...

Page 17: ...cap There is a risk of heating the air jet cap is the nozzles are not regularly cleaned Layers of soap or deposit can hinder the circulation of the air To clean the nozzles please follow the next stag...

Page 18: ...he appliance should be recorded in the service logbook provided and should be retained by the owner of the appliance for reference Do not use the product if the service logbook has not been signed aft...

Page 19: ...MAINTENACE FORM LOGBOOK Service providers signature Clients signature Parts fitted Service providers Comments Service or repair undertaken Date 18 18...

Page 20: ...ulle Tel 33 05 46 42 04 16 Fax 33 05 46 43 06 38 www groupe reval com UNITED KINGDOM Reval Continuing Care Limited Unit 17 Brynmenyn Busines Centre St Theodore s Way Brynmenyn CF32 9TZ BRIDGEND WALES...