Packaging, Transport and Installation
27
Temporary storage:
Also in case of an interim storage the device must be stored dry and within the specified
ambient temperature range.
4.4 Conditions for the Installation Site
NOTICE
N6.0021
Ambient temperature
Temperatures outside the permitted range
−
Electronic and mechanical components may be damaged.
−
The performance data alters to an unknown extent.
•
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
−
Installation height: max. 2
000 m above sea level
−
Ambient temperature: 5 °C
– 40 °C
−
Maximum relative humidity < 80 % (at ambient temperatures
≤ 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
=
–(U
T
– 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N7.0015
Humidity
High relative humidity
−
Electronic and mechanical components may be damaged.
−
The performance data alters to an unknown extent.
•
The relative humidity in the vicinity of the device should be kept as
low as possible.