First Commissioning
23
NOTICE
To clamp a maximum of five test sieves and a collecting pan, shorter threaded rods
are available for the sieve clamping unit "standard". For sieving processes with only one to three
test sieves, the shorter threaded rods should be used. Long, projecting threaded rods disturb
the spreading of the sample material due to their natural vibration behaviour.
Pos : 13.9 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 Siebs pannei nheit "com fort" @ 14\m od_1455695607243_9.docx @ 98656 @ 2 @ 1
6.2 Sieve Clamping Unit "comfort"
Pos : 13.10 / 0005 R ET SCH/ 0005 Bedienungsanl eitungen/ AS 200 / AS 300 s erie/Modul e/ 2010 Siebs pa nnei nheit "c omf ort" AS 300 c ontrol B @ 18\mod_1473658606761_9. docx @ 128424 @ @ 1
Put both quick clamping units (
F
) on a flat surface with the green quick clamping lever (
F1
)
facing down.
Place the clamping lid (
D
) with the top side (plane side) face down on the quick clamping
units (
F
).
Place the O-ring (
OR
) on the cone shaped assembly aid (
MH2
) and slide it down into the
designated groove.
Put the assembly aid (
MH2
) in the opening of the clamping lid (
D
) in such a way that the
cone shaped tip is sticking out.
Place the assembly aid ring (
MH1
) on the assembly aid (
MH2
) and slide it down. This
presses the O-ring on the quick clamping unit and fixes the clamping lid.
Repeat this procedure for the other side.
Fig. 7:
Assembly of the clamping lid
Screw one hexagonal nut (
G
) on the thread of each of the support rods (
E
).
Screw both support rods (
E
) into the designated threaded holes (
SB
) in the sieve plate (
ST
)
and lock them with the hexagonal nuts (
G
).
Firmly tighten the hexagonal nuts (
G
) by means of a 19 mm open-end wrench.
MH1
MH2
OR
F
D
F1
F2