7-27
7. Alarms
Item Alarm
Priority
Definition
Action needed
45
Technical
Fault (3005)
High
Protecting
module Comm
Stop.
Contact your service
providers.
46
Technical
Fault (3006)
High
Power Board
Comm Stop.
Contact your service
providers.
47
Technical
Fault (3007)
High
Key Board
Comm Stop.
Contact your service
providers.
48
Technical
Fault (3008)
High
Uart extended
module Comm
Stop.
Contact your service
providers.
49
Battery1
recharge
failed,
temperature
low
Low
Battery 1 temp.
error. Unable to
charge.
Connect to the
external power
supply.
Wait for the machine
to return to normal
temperature.
50
Battery2
recharge
failed,
temperature
low
Low
Battery 2 temp.
error. Unable to
charge.
Connect to the
external power
supply.
Wait for the machine
to return to normal
temperature.
51
CO
2
sampleline
occluded
Low
The sampling lin
e has an error or
is occluded.
Replace the
sampling line.
52
CO
2
sensor
overheated
Low
The CO
2
sensor
assembly temp
erature is too hi
gh (greater than
50℃ ).
Replace the
CO
2
module.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...