7-19
7. Alarms
ventilation in VCV mode normally.
2. Set the VT low alarm limit to be greater than the
current VTe.
3. Verify that the [
Low Tidal Volume
] alarm is activated.
7.12.11 Low Minute Volume
1. Make sure a demonstration lung is connected to
the ventilator and start ventilation in VCV mode
normally.
2. Set the MV low alarm limit to be greater than the
current MV.
3. Verify that the Low Tidal Volume alarm is activated.
7.13 Alarm Troubleshooting Table
The below table is a list of the alarm messages
displayed by the ventilator, along with their definitions
and suggested corrective actions. These corrective
actions are sequenced to correct the most probable
malfunction or to present the most efficient corrective
action first. The proposed actions, however, may not
always correct the particular problem.
This chapter lists physiological and technical alarm
messages.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...