Chapter 5
Repair
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
5-15
b.
Verify that the oxygen source is disconnected from the ventilator.
c.
Verify that
Oxygen Supply
reads
OFF
.
d.
Return the hardware screen to its initialized state.
Figure 5-10: Diagnostic Mode - Hardware Screen (Safety Solenoid Energized)
Oxygen Flow Sensor
1. Connect the oxygen source to the ventilator.