Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-33
9. To remove the backlight inverter PCB, remove the two Phillips head
screws that hold it to the GUI.
Figure 9-35: Backlight Inverter PCB and VGA Display (10.4-in. GUI)
Transition PCBA, LCD,
IR Touch Frame, and
Backlight Inverter
PCB
(2nd Generation
10.4-in. GUI)
Follow these steps to remove the backlight inverter PCB, LCD, and touch frame
(Figure 9-36). Reverse removal steps to install.
1. Remove the 2.5-mm screws securing the GUI to the enclosure.
2. Remove the rotary encoder (see “Rotary Encoder (10.4-in. GUI’s)” on
page 9-28). The rotary encoder has the only harness connector not
accessible within the GUI assembly.
3. Gently pry the GUI assembly from the ventilator enclosure enough to
access and disconnect the harness connectors.
4. Remove the rotary encoder from the GUI assembly.
5. Disconnect harnesses and cables from the touch frame, transition
PCBA, backlight inverter PCB, and keyboard interconnects.
6. Set the GUI face down on a nonabrasive surface.
10.4-in. VGA display
Backlight
inverter PCB