Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
5-13
5.8 Generating a
Diagnostic Report
(DRPT)
1. With the BiPAP Focus ventilator in Diagnostic mode and the
Diagnostics
end of the serial communications cable connected to the
ventilator, start HyperTerminal, open the BiPAP Focus
Communications file, or click on the BiPAP Focus Communications
icon on the PC (if created).
2. In the HyperTerminal window menu, select
Transfer > Capture Text...
.
3. In the HyperTerminal
Capture Text
window, click
Browse
.
4. In the HyperTerminal
Select Capture File
window, select where to store
the capture file and enter a filename. Leave
Save as type
as TXT
(generic text file).
5. In the HyperTerminal
Select Capture File
window, click
Save
.
6. In the HyperTerminal
Capture Text
window, click
Start
.
7. Type
DRPT
(all caps) in the dialog box, and press the
Enter
key. This
retrieves information from the significant event log.
8. If the ventilator is connected and communication occurs, a response
similar to the following appears:
Respironics Inc. BiPAP Focus Ventilator
Serial Number: 8772060206-04
Date & Time: 6/19/09 20:49
Event ID
Date
Time
905
Pwr On reset
6/19/09
20:43
300
POST Successful
6/19/09
20:43
102
Power Off
6/19/09
20:42
437
Battery Voltage Failed
6/19/09
20:42
V 615,615
639
Disconnect
6/19/09
20:42
635
Battery Depleted
6/19/09
20:42
637,633
644
Apnea
6/19/09
20:42
639
Disconnect
6/19/09
20:
635
Battery Depleted
6/19/09
20:41
687,649
643
Low Battery
6/19/09
20:41
714,723
905
Pwr On reset
6/19/09
20:41
300
POST Successful
6/19/09
20:41
102
Power Off
6/19/09
20:41
905
Pwr On reset
6/19/09
20:26
300
POST Successful
6/19/09
20:26
102
Power Off
6/14/09
16:52