Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
7-13
7.6.2
Performing the System Calibration
The calibration procedure provides instructions for the remote calibration of
the BiPAP Focus. The results of this calibration must be printed to a local
printer.
NOTE:
The real-time clock calibration (section 7.5) must be complete before
attempting the system calibration procedure.
Required equipment:
•
Utility Software, Respironics Utility Tools
(
http://my.respironics.com
and click on
BiPAP Focus Calibration
Software)
•
Pneumatic calibration analyzer (P/N 1040310 or equivalent) capable
of measuring low pressure (cmH
2
O) and flow rate (LPM).
•
Flow source capable of providing up to 120 LPM of air to provide
negative/reverse flow
•
Windows-compatible personal computer running Windows XP version
software.
•
A working default printer connected to the computer before beginning
the calibration procedure.
•
BiPAP Focus communications cable kit (P/N 1039800)
•
Flow control valve (P/N 1006120)
•
Oxygen enrichment adapter (P/N 312710))
•
Pressure tubing
•
Smooth-bore patient tubing
NOTE:
NOTE:
Serial port must be set to COM1.
NOTE:
If not already done, perform the ventilator run-in (section 7.3).
1. Record the serial number and model number
exactly
as they appear on
the bottom of the ventilator.
2. Connect the serial communications cable between the service PC and
the ventilator, with the software/RASP end connected to the ventilator.
3. Ensure that the ventilator is connected to AC power, then power up in
Diagnostic mode.