Appendix B
1062788
BiPAP Vision Service Manual © Respironics, Inc.
B-5
B.10 Settings
NOTE:
See the BiPAP Vision
Clinical Manual
for PAV/T information.
Parameter
Range
Increments
IPAP
4 to 40 cmH
2
O
1 cmH
2
O
EPAP
4 to 20 cmH
2
O
1 cmH
2
O
CPAP
4 to 20 cmH
2
O
1 cmH
2
O
Rate
4 to 40 BPM
1 BPM
Timed inspiration
0.5 to 3.0 s
0.1 s
IPAP rise time
0.05 to 0.4 s
0.05, 0.1, 0.2,
or 0.4 s
Oxygen concentration (%O
2
)
(if oxygen module is installed)
21 to 100%
21 to 25%: 4%
25 to 100%: 5%
Alarm
Range
Increments
High pressure
5 to 50 cmH
2
O
1 cmH
2
O
Low pressure
0 (disabled) to 40
cmH
2
O
1 cmH
2
O
Low pressure delay
0 to 60 s
1 s
Apnea
0 (disabled), 20 to
40 s
0 (disabled), 20,
30, 40 s
Low minute ventilation
(if alarm module is installed)
0 (disabled) to 99
L/min
1 L/min
High rate
(if alarm module is installed)
4 to 120 BPM
1 BPM
Low rate
(if alarm module is installed)
4 to 120 BPM
1 BPM