Appendix B
B-2
BiPAP Vision Service Manual © Respironics, Inc.
1062788
B.4 Pressure
Protection against
electrical shock
Class I
Type
BF
Degree of protection
against harmful ingress of
water
Ordinary equipment, IPX0
Electromagnetic
compatibility
Meets requirements of IEC 601-1-2
Earth resistance
< 0.10 ohms
Earth leakage current
Normal pole, no earth, L2: < 300 µA
Reverse pole, no earth, L2: < 300 µA
Reverse pole, no earth, no L2:
< 1000 µA
Normal pole, no earth, no L2:
< 1000 µA
Insulation resistance
> 2 megaOhms
Alarm sound
70 to 85 dBA peak at a distance of 1 meter
Noise
No specification: various test instruments,
test procedures, and device operating
conditions produce varying results.
Output
4 to 40 cmH
2
O
Dynamic regulation
+ 2 cmH
2
O at sinusoidal flow at + 100 L/min
Static regulation
+ 2 cmH
2
O from -60 to 120 L/min
Elevation
0 to 5000 ft. above sea level